3-3 Environmental Factor Setting ⇒
To set different stability factors according to different environment.
The worse the environment is, the larger stability factor is.
The larger value, the faster it will stable but less accurate.
Stability factor is set for "05".
Do not set for Hold Function mode (F8).
Please make sure MINI JUMPER of SWA1 is connected to ADJ on PC board before
setting. Please make sure MINI JUMPER of SWA1 is connected to LOCK on PC board
after setting.
Press
or
Press
key, the
display shows default setting
Press
key,
input the stable factor
Press
key
key to select the F2 function⇒the display shows
key increases the flash value
by one. (input 01~15 digits)