Excell SI-132 Service Manual page 44

High resolution weighing scale
Hide thumbs Also See for SI-132:
Table of Contents

Advertisement

EXCELL PRECISION CO., LTD
Ex3: 30kg (5kg, 5kg ,10kg, 10kg)
Display
Key Press
03 CLn
TARE
L0
TARE
L1 1
TARE
L2 1
TARE
L3 2
TARE
L4 2
TARE
L5 1
NET/GROSS
03 CLn
Ex4: 30kg (1kg, 2kg ,5kg, 10kg, 2kg, 10kg)
Display
Key Press
03 CLn
TARE
L0
TARE
L1 1
TARE
L2 2
TARE
L3 5
TARE
L4 A
TARE
L5 2
TARE
L6 A
TARE
L7 1
NET/GROSS
03 CLn
Notes:
In the process of
In the process of
finish and save the 2, 3, 4, 5, 6, 7, or 8 points calibration.
In the process of
02005302
Note
Into linearity calibration
First point (zero), remove weigh pan and press the Tare key
Put 5kg on and press the Tare key
Put 5kg on and press the Tare key
Put 10kg on and press the Tare key
Put 10kg on and press the Tare key
To finish linearity adjustment (5 points linearity calibration)
Note
Into linearity calibration
First point (zero), remove weigh pan and press the Tare key
Put 1kg on and press the Tare key
Put 2kg on and press the Tare key
Put 5kg on and press the Tare key
Put 10kg on and press the Tare key
Put 2kg on and press the Tare key
Put 10kg on and press the Tare key
To finish linearity calibration (7 points linearity calibration)
,
, press
NET/GROSS
,
,
,
,
, press
key to finish and save the 9 points calibration.
TARE
43
key to abort the linearity calibration.
,
, or
press
key to
NET/GROSS
ZSME300000237