Endress+Hauser Proline 500 Brief Operating Instructions
Endress+Hauser Proline 500 Brief Operating Instructions

Endress+Hauser Proline 500 Brief Operating Instructions

Modbus rs485 transmitter with ultrasonic time-of-flight sensor

Advertisement

Quick Links

KA01476D/06/EN/02.24-00
71636525
2024-01-31
Products
Brief Operating Instructions
Proline 500
Modbus RS485 transmitter
with ultrasonic time-of-flight sensor
These instructions are Brief Operating Instructions; they are
not a substitute for the Operating Instructions pertaining to
the device.
Brief Operating Instructions part 2 of 2: Transmitter
Contain information about the transmitter.
Brief Operating Instructions part 1 of 2: Sensor →  3
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline 500

  • Page 1 Solutions Services KA01476D/06/EN/02.24-00 71636525 2024-01-31 Brief Operating Instructions Proline 500 Modbus RS485 transmitter with ultrasonic time-of-flight sensor These instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device. Brief Operating Instructions part 2 of 2: Transmitter Contain information about the transmitter.
  • Page 2 Proline 500 Modbus RS485 Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
  • Page 3 These Brief Operating Instructions are Brief Operating Instructions Part 2: Transmitter. The "Brief Operating Instructions Part 1: Sensor" are available via: • Internet: www.endress.com/deviceviewer • Smart phone/tablet: Endress+Hauser Operations App Detailed information about the device can be found in the Operating Instructions and the other documentation: • Internet: www.endress.com/deviceviewer •...
  • Page 4: Table Of Contents

    Table of contents Proline 500 Modbus RS485 Table of contents About this document ............. . 5 Symbols used .
  • Page 5: About This Document

    Proline 500 Modbus RS485 About this document About this document Symbols used 1.1.1 Safety symbols DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury.
  • Page 6 About this document Proline 500 Modbus RS485 Symbol Meaning Potential equalization connection (PE: protective earth) Ground terminals that must be connected to ground prior to establishing any other connections. The ground terminals are located on the interior and exterior of the device: •...
  • Page 7: Safety Instructions

    Proline 500 Modbus RS485 Safety instructions Safety instructions Requirements for the personnel The personnel must fulfill the following requirements for its tasks: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task. ‣ Are authorized by the plant owner/operator.
  • Page 8: Workplace Safety

    Safety instructions Proline 500 Modbus RS485 Workplace safety When working on and with the device: ‣ Wear the required personal protective equipment as per national regulations. Operational safety Damage to the device! ‣ Operate the device in proper technical condition and fail-safe condition only.
  • Page 9: Product Description

    Proline 500 Modbus RS485 Product description Product description The measuring system consists of a transmitter and two or one sensor sets. The transmitter and sensor are mounted in physically separate locations. They are interconnected via sensor cable(s). A0041373 Transmitter with integrated ISEM...
  • Page 10 Mounting procedure Proline 500 Modbus RS485 The transmitter can be mounted in the following ways: • Post mounting • Wall mounting Wall mounting Required tools Drill with drill bit ⌀ 6.0 mm 18 (0.71) 10 (0.39) 20 (0.79) 100 (3.94) A0029068 ...
  • Page 11 Proline 500 Modbus RS485 Mounting procedure Pipe mounting Required tools Open-ended wrench AF 13 WARNING Order code for "Transmitter housing", option L "Cast, stainless": cast transmitters are very heavy. They are unstable if they are not mounted on a secure, fixed post.
  • Page 12 Mounting procedure Proline 500 Modbus RS485 4.2.2 Turning the transmitter housing To provide easier access to the connection compartment or display module, the transmitter housing can be turned. – – 3 mm 4 mm 7 Nm (5.2 lbf ft) A0029993 ...
  • Page 13 Proline 500 Modbus RS485 Mounting procedure 4 mm 7 Nm (5.2 lbf ) A0043150  4 Ex housing Loosen the fixing screws. Turn the housing to the desired position. Tighten the securing screws. Endress+Hauser...
  • Page 14 Mounting procedure Proline 500 Modbus RS485 4.2.3 Turning the display module The display module can be turned to optimize display readability and operability. – – – 3 mm A0030035 Depending on the device version: Loosen the securing clamp of the connection compartment cover.
  • Page 15 Proline 500 Modbus RS485 Mounting procedure 4.2.4 Transmitter post-installation check The post-installation check must always be performed after the following tasks: • Mounting the transmitter housing: • Post mounting • Wall mounting • Turning the transmitter housing • Turning the display module Is the device undamaged (visual inspection)? ...
  • Page 16: Cover Locking

    Mounting procedure Proline 500 Modbus RS485 Cover locking NOTICE Order code "Transmitter housing", option L "Cast, stainless": The covers of the transmitter housing are provided with a borehole to lock the cover. The cover can be locked using screws and a chain or cable provided by the customer on site.
  • Page 17: Weather Protection Cover

    Proline 500 Modbus RS485 Mounting procedure Weather protection cover 280 (11.0) 255 (10.0) 146 (5.75) 134 (5.3) 12 (0.47) 30 (1.18) A0029553  5 Engineering unit mm (in) Endress+Hauser...
  • Page 18: Electrical Connection

    Electrical connection Proline 500 Modbus RS485 Electrical connection WARNING Live parts! Incorrect work performed on the electrical connections can result in an electric shock. ‣ Set up a disconnecting device (switch or power-circuit breaker) to easily disconnect the device from the supply voltage.
  • Page 19 Standard installation cable is sufficient 5.2.3 Connecting cable between the transmitter and sensor Sensor cable for sensor - transmitter: Proline 500 Standard cable • TPE: -40 to +80 °C (-40 to +176 °F) • TPE armored: -40 to +80 °C (-40 to +176 °F) •...
  • Page 20 Electrical connection Proline 500 Modbus RS485 5.2.4 Terminal assignment Transmitter: supply voltage, input/outputs The terminal assignment of the inputs and outputs depends on the individual order version of the device. The device-specific terminal assignment is documented on an adhesive label in the terminal cover.
  • Page 21: Connecting The Measuring Device

    Proline 500 Modbus RS485 Electrical connection Connecting the measuring device NOTICE An incorrect connection compromises electrical safety! ‣ Only properly trained specialist staff may perform electrical connection work. ‣ Observe applicable federal/national installation codes and regulations. ‣ Comply with local workplace safety regulations.
  • Page 22 Electrical connection Proline 500 Modbus RS485 Connecting the sensor cable to the transmitter A0044340 Loosen the securing clamp of the connection compartment cover. Unscrew the connection compartment cover. Route the two sensor cables of channel 1 through the slackened top union nut of the cable entry.
  • Page 23 Proline 500 Modbus RS485 Electrical connection 11. Screw on the connection compartment cover. 12. Tighten the securing clamp of the connection compartment cover. 13. After connecting the sensor cable(s): Connect the signal cable and the supply voltage cable →  24.
  • Page 24 Electrical connection Proline 500 Modbus RS485 5.3.2 Connecting the signal cable and the supply voltage cable A0026781 Terminal connection for supply voltage Terminal connection for signal transmission, input/output Terminal connection for signal transmission, input/output or terminal connection for network connection via service interface (CDI-RJ45; non-Ex) Protective earth (PE) –...
  • Page 25 Proline 500 Modbus RS485 Electrical connection N ic N ic t e r t e r e r g ö f f e r g ö f f iz e iz e t e n t e n r ir...
  • Page 26 Electrical connection Proline 500 Modbus RS485 22 mm 24 mm – A0029816 10. Connect the cable according to the terminal assignment.  Signal cable terminal assignment: The device-specific terminal assignment is documented on an adhesive label in the terminal cover.
  • Page 27 Proline 500 Modbus RS485 Electrical connection 5.3.3 Integrating the transmitter into a network This section only presents the basic options for integrating the device into a network. Integrating via the service interface The device is integrated via the connection to the service interface (CDI-RJ45).
  • Page 28: Ensuring Potential Equalization

    Electrical connection Proline 500 Modbus RS485 Ensuring potential equalization 5.4.1 Requirements No special measures for potential equalization are required. For devices intended for use in hazardous locations, please observe the guidelines in the Ex documentation (XA). Endress+Hauser...
  • Page 29: Hardware Settings

    Proline 500 Modbus RS485 Electrical connection Hardware settings 5.5.1 Setting the device address The device address must always be configured for a Modbus slave. The valid device addresses are in the range from 1 to 247. Each address may only be assigned once in a Modbus RS485 network.
  • Page 30: Ensuring The Degree Of Protection

    Electrical connection Proline 500 Modbus RS485 5.5.2 Activating the terminating resistor To avoid incorrect communication transmission caused by impedance mismatch, terminate the Modbus RS485 cable correctly at the start and end of the bus segment. ‣ A0029632 Switch DIP switch no. 3 to On.
  • Page 31: Post-Connection Check

    Proline 500 Modbus RS485 Electrical connection Post-connection check Are cables or the device undamaged (visual inspection)?  Is the protective earthing established correctly?  Do the cables used comply with the requirements ?  Are the mounted cables relieved of tension? ...
  • Page 32: Operation Options

    Operation options Proline 500 Modbus RS485 Operation options Overview of operation options A0030213 Local operation via display module Computer with Web browser (e.g. Internet Explorer) or with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) Mobile handheld terminal with SmartBlue App Control system (e.g.
  • Page 33: Structure And Function Of The Operating Menu

    Proline 500 Modbus RS485 Operation options Structure and function of the operating menu 6.2.1 Structure of the operating menu Operating menu for operators and maintenances Language Operation task-oriented Setup Diagnostics Operating menu for experts Expert function-oriented A0014058-EN  6 Schematic structure of the operating menu 6.2.2...
  • Page 34: Access To Operating Menu Via Local Display

    Operation options Proline 500 Modbus RS485 Access to operating menu via local display X X X X X X 19.184 mA 12.5 X X X X X X Language 20.50 Ã English Deutsch Español Français Xxxxxxx A B C D E F G H I J K M N O P Q R S T U V W +0.000 Xx...
  • Page 35 Proline 500 Modbus RS485 Operation options 6.3.1 Operational display Explanatory symbols for the measured value Status area • Depends on the device version, e.g.: The following symbols appear in the status area of the • : Volume flow operational display at the top right: •...
  • Page 36 Operation options Proline 500 Modbus RS485 Numeric editor Confirms selection. Moves the input position one position to the left. Exits the input without applying the changes. Inserts decimal separator at the cursor position. Inserts minus sign at the cursor position.
  • Page 37: Access To The Operating Menu Via The Operating Tool

    Proline 500 Modbus RS485 Operation options Operating key Meaning Escape key combination (press keys simultaneously) In menu, submenu • Pressing the key briefly: • Exits the current menu level and takes you to the next higher level. • If help text is open, closes the help text of the parameter.
  • Page 38: System Integration

    System integration Proline 500 Modbus RS485 System integration For detailed information on system integration, see the Operating Instructions for the device →  3 • Overview of device description files: • Current version data for the device • Operating tools •...
  • Page 39: Configuring The Measuring Device

    Proline 500 Modbus RS485 Commissioning X X X X X X X 20.50 Main menu 0104-1 Display language English Operation Setup Display language 0104-1 English à Deutsch Español Français Display language 0104-1 à English Deutsch Español Français Hauptmenü 0104-1 Sprache...
  • Page 40: Protecting Settings From Unauthorized Access

    Diagnostic information Proline 500 Modbus RS485 Example: Available submenus, wizards Meaning Relay output Double pulse output Display Configuration of the display format on the local display Low flow cut off Configuration of the low flow cut off Advanced setup Additional parameters for configuration: •...
  • Page 41 Proline 500 Modbus RS485 Diagnostic information X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage Diagnostics 2 Diagnostics 3 Supply voltage (ID:203) S801 0d00h02m25s...
  • Page 44 *71636525* 71636525 www.addresses.endress.com...

Table of Contents