Page 1
ESTUFA DE DESECACIÓN - INCUBACIÓN OVENS AND INCUBATORS ÉTUVES DE DESSÈCHEMENT - CULTURE SERIES - SERIES - SÉRIES 631,632, 636 Este manual es parte inseparable del aparato por lo que debe estar disponible a todos los usuarios del equipo. Le recomendamos leer atentamente el presente manual y seguir rigurosamente los procedimientos de uso para obtener las máximas prestaciones y una mayor duración del mismo.
Page 2
EQUIPMENT. Please bear in mind the following: This manual is inseparable from the Nahita ovens and incubators, so it should be available for all the users of these equipments. Follow all required applicable laws about safety and health (electrical security devices, inflammable, toxic and/or pathological products, etc.).
Page 3
4. INSTALLATION / SETTING UP ....14 1. USES OF THE APPARATUS Nahita ovens and incubators are very useful in research, hospital or industrial laboratories for drying samples, disinfection of material and maintenance of cell and microorganisms cultures. Made of furnace-painted steel, they present a easy-to-clean stainless steel inner chamber. To control sample evolution they are provided with a double glass window.
ENGLISH Parts of the equipment 1.1 Control panel lation is continuous. 1.2 On/off switch 1.6 On the Series 631 Plus there is a button to 1.3 Ventilation grilles The air outlet is the chim- select one of two heating speeds); In speed 1, a ney located at the top voltage of 110V enters the resistance, while in 1.4 Socket...
Page 5
ENGLISH Series 632 Plus - Forced air drying ovens Code JBD005 JBD006 JBD007 JBD008 JBD010 Capacity 30 L 45 L 65 L 125 L 230 L Temp. range 80-250ºC (recommended) Resolution 1 ºC (configurable to 0.1ºC) * Temp. fluctuation ± 1.0 ºC Homogeneity Consumo 0.9 Kw...
ENGLISH Please check that all the accessories are enclosed with the equipment: - 2 trays and 4 guides already placed. - Shuko wire - User’s manual - 1 spare fuse, amps depending on capacity. Installation Before using this instrument, it is convenient for you to familiarize with its components and essentials. PLEASE READ THOROUGHLY THE INSTRUCTIONS BEFORE CONNECTING AND OPERATING WITH THIS EQUIPMENT.
Page 7
ENGLISH selected temperature in 10 ºC for ovens (series 631 Plus and 632 Plus) or in 5 ºC for incubators (series 636 Plus). Let the equipment cool down and set it again. Working time adjustment By default, whenever turning the equipment on the working time is off, so the equipment will operate until turning it off by pressing on/off switch (1.2).
ENGLISH Thin adjustment of inner parameter SC To adjust real temperature (measured with an external thermometer) and temperature displayed by the equipment, it is possible to make a thin adjustment through parameter SC: Keep pressed “FUNC” key until ‘LOCY’ command appears at the upper display. Press again ”FUNC”...
Page 9
ENGLISH that usually handle the equipment (burns, hurts…) or damages in nearby equipments. In the event of breakdown please contact your distributor to overhaul through Nahita Technical Assistance Department. Cleaning Never use scourers or substances that can grate for cleaning metallic parts such as stainless steel, aluminium, coatings, etc.
ENGLISH 7. INTERNAL PARAMETERS To access the parameters and with the stove on, keep the FUNC key pressed until “LOCY” is displa- yed. Enter password 8000 and FUNC key to confirm access INTERNAL PARAMETERS P=20 These are parameters that are modified automatically I=100 at the time of auto-tuning.