Summary of Contents for Nahita GLC001

  • Page 1 For safety reasons, please read the operating manual carefully before starting the operation. Components subjected to heat, humidity, and biological changes can lead to functional degradation, and even loss of safety, leading to accidents! User's Manual of GLC001 Version 1, July 2023...
  • Page 2 * The maintenance of the instrument can only be done byMagal or authorized dealers ofMagal.; * Please NEVER use the following materials in the centrifuge: -Inflammable and explosive materials; -Strengthen chemical materials; User's Manual of GLC001 - 1 - Version 1, July 2023...
  • Page 3 The consequences of ignoring safety procedures Any action ignoring the safety procedure, the law and rules, or other relative rules will cause damage and danger to the person, device and environment. User's Manual of GLC001 - 2 - Version 1, July 2023...
  • Page 4: Table Of Contents

    7.1 Emergency door opening cover………………………….……………… 21 7.2 Fault alarm information…………..……………………………………… 21 7.3 Machine operation records…………………….………………………… 24 8. technical data ……………………………..……………………………………25 9. packing list ……………………..……………………………………………26 10. certificate of compliance …...…………………………………………………27 User's Manual of GLC001 - 3 - Version 1, July 2023...
  • Page 5 30 cm around the centrifuge, no hazardous substances or items blocking vents within 30 cm around the centrifuge. If in the use of Magal low-speed BenchtopAir-refrigerate Centrifuge, the User's Manual of GLC001 - 4 - Version 1, July 2023...
  • Page 6 ●After using for a period of time, the provisions of maintenance should be strictly in accordance with the "cleansing and disinfection," 1.2 The relevant quoting standards ● GB4793.1-2007 Part I of Safety Requirements of Measurement, control of User's Manual of GLC001 - 5 - Version 1, July 2023...
  • Page 7 & special requirements of laboratories electrical centrifuge; ● GB191-2000 Icon of packaging & Transportation; ● GB6587.6-86 Transport-Test of Electronic Measurement Instruments ● GB/T 14710-1993 Environmental requirement & test way of medical electrical equipment; User's Manual of GLC001 - 6 - Version 1, July 2023...
  • Page 8: Centrifuge

    English 2. Introduction to GLC001 Low Speed Centrifuge 2.1 Introduction to the Appearance of GLC001 Low Speed Centrifuge Figure 1: Front Profile of GLC001 Low Speed Centrifuge Figure 2: Rear Appearance of GLC001 Low Speed Centrifuge User's Manual of GLC001...
  • Page 9: Overview

    6. Power socket and power switch 7. Emergency door opening 2.2 overview The GLC001 low-speed centrifuge is a low-speed centrifuge used for routine analysis in medical laboratories, biochemical and molecular biology research, and industrial laboratories. It can be widely used in clinical medicine, biology, chemistry, genetic engineering, immunology, and other fields.
  • Page 10: Security Protection Agency

    2.4 Security Protection Agency The GLC001 low-speed centrifuge has a series of safety protection mechanisms: The frame and protective steel ring are made of steel plates, and the internal cavity is User's Manual of GLC001...
  • Page 11: Machine Placement Requirements

    During the operation of the rotor, it is strictly prohibited to manually power off, otherwise it may cause damage to the control circuit! User's Manual of GLC001 - 10 - Version 1, July 2023...
  • Page 12: Optional Accessories

    English 3. Optional accessories The rotors of various specifications of GLC001 low-speed centrifuge are provided for users. You can purchase the GLC001 low-speed centrifuge according to your actual use requirements (see "Table 1: Rotor Types and Technical Parameters" for details).
  • Page 13: Preparation Before Use

    4.2 Choose a reasonable placement location The GLC001 low-speed centrifuge can only be used indoors, and its placement should meet the following requirements: When the centrifuge is running, a safe distance of 20cm must be kept around. No harmful Dangerous goods shall be placed within this safe distance, and relevant personnel shall not stay;...
  • Page 14: Fixed Placement Machine

    Figure 4. Introduction to control panel: 1. set selection key 2. volume control key 3. centrifugation time setting key 4. speed / relative centrifugal force setting key, switch key User's Manual of GLC001 - 13 - Version 1, July 2023...
  • Page 15: Power On

    The bottom of the door hook will contact the travel switch and the door has been locked. When closing the door, press the door cover properly User's Manual of GLC001 - 14 - Version 1, July 2023...
  • Page 16: Installing The Rotor

    After installing the rotor, before each use or after a period of use, check whether the installation position of the rotor changes. If necessary, tighten the locking sleeve again to ensure that the rotor is firmly installed. User's Manual of GLC001 - 15 - Version 1, July 2023...
  • Page 17 English Figure 5: schematic diagram of installing rotor body User's Manual of GLC001 - 16 - Version 1, July 2023...
  • Page 18: Calculation Of Rotor Load

    5.6 Calculation of rotor load ● Calculation of maximum load When the GLC001 low-speed centrifuge runs at low speed, there is a huge centrifugal force; When designing each rotor, it is required to have sufficient mechanical strength when working at the maximum rated speed - that is, "safety factor";...
  • Page 19: Safe Use Of Rotors

    For example, the following parameters need to be set: root Speed r/min. Time min. 4003 4000 5.9.2 Rotor number setting: Press the key on the control panel User's Manual of GLC001 - 18 - Version 1, July 2023...
  • Page 20 3. Note: When DEC is set to 0, the shutdown is free parking, and the system has no brake intervention! After setting, it is shown in the following figure: User's Manual of GLC001 - 19 - Version 1, July 2023...
  • Page 21: Parameter Settings Other

    5.10.6 After the machine speed stabilizes, if necessary, the parameters such as speed/centrifugal force, time, and lifting speed can be modified again. After resetting User's Manual of GLC001 - 20 - Version 1, July 2023...
  • Page 22 Note: "The maximum centrifugal force value is related to the maximum centrifugal radius"; The "centrifugal force value" set should take into account the radius of the rotor and the shape of the centrifugal container. User's Manual of GLC001 - 21 - Version 1, July 2023...
  • Page 23: Maintenance Matters

    6.1.2 Implementation of steam sterilization The lifespan of the attachment is related to the frequency and use of sterilization. User's Manual of GLC001 - 22 - Version 1, July 2023...
  • Page 24 6.1.3.5 Use a cloth or tweezers to remove debris from the centrifuge chamber. 6.1.3.6 The connecting parts of the motor shaft and rotor shaft hole should be coated with lubricating grease. 6.1.3.7 Steps for Maintaining the Motor Spindle: User's Manual of GLC001 - 23 - Version 1, July 2023...
  • Page 25 This machine is a precision instrument and equipment. During transportation and storage, please pay attention to moisture and shock resistance, and do not lay it horizontally or upside down. User's Manual of GLC001 - 24 - Version 1, July 2023...
  • Page 26: Warranty (Repair) Basis

    Attention: This method is only allowed to be used in emergency situations and must not be used casually! User's Manual of GLC001 - 25 - Version 1, July 2023...
  • Page 27: Fault Alarm Information

    After an abnormality occurs, the power should be turned off first and restarted after the fault is resolved. User's Manual of GLC001 - 26 - Version 1, July 2023...
  • Page 28 Check if the cable between the display board COMMUNICATION ● Erro8 and the driver board is properly connected. ERROR External input voltage too high or acceleration ● Erro9 overvoltage too fast User's Manual of GLC001 - 27 - Version 1, July 2023...
  • Page 29 2. The PCB board and power socket fuse are broken, please have a professional on, the panel replace the fuse. Nixie tube 3. If it cannot be ruled out, please contact our company's maintenance indicator does personnel. not light up User's Manual of GLC001 - 28 - Version 1, July 2023...
  • Page 30 4. Rotate the transmission shaft by hand. If the transmission cannot be smooth, smoke odor please ask professional personnel from the manufacturing unit to solve the problem. Exceeding the above fault 1. Solved by professional personnel from the manufacturing unit. content User's Manual of GLC001 - 29 - Version 1, July 2023...
  • Page 31: Machine Operation Records

    (one for each machine). The recommended operation record table is as follows: Single tube machine weight test sample rotor Work (including Speed time operator date name number situatio samples) User's Manual of GLC001 - 30 - Version 1, July 2023...
  • Page 32 ● 30 ㎏ Net weight ● GB4793.1-2007《测量、控制实验室用电气设备的安全要求第 Referenced standards 一部分:通用安全要求》 ; ● GB4793.7-2001《测量、控制实验室用电气设备的安全要求 实 验室用离心机的特殊要求》 ; ● GB191-2000《包装储运图示标志》 ; ● GB6587.6-86《电子测量仪器运输试验》 ; ● GB/T14710-1993《医用电气设备环境要求及试验方法》 ; ● 企业标准: 《低速离心机》 。 User's Manual of GLC001 - 31 - Version 1, July 2023...
  • Page 33: Packing List

    English 9. packing list GLC001 Low Speed Centrifuge packing list name Serial number notes Number main engine national Power cord standard Turning head/hanging basket Supply according to the order contract Special Hex key 20ml lubricating oil Certificate of Conformity Operation, Maintenance and Repair...