Technical specifications

10.7 Mechanical requirements

Transport and storage (packed)
Temperature range
Temperature gradient
Humidity
Air pressure
10.7 Mechanical requirements
In operation
Vibrations
Shock
Packaged unit
Vibrations
Shock
Falling / tipping over

10.8 Safety regulations

This product has been assigned a CE marking in compliance with the directives of guideline
93/42/EEC of June 14, 1993, concerning medical products.
Safety regulations
Safety standards
Protection class
Degree of protection
Conformity
34
-20 °C ... +55 °C ambient temperature
Max. 5 °C/h, no condensation
10 ... 80 %, non-condensing
200 ... 1050 hPa
According to EN 60068-2-6 and EN 60721-3-3
According to EN 60068-2-27 and EN 60721-3-3
Class 2M2 according to EN 60068-2-64 and
EN 60721-3-2
Class 2M2 according to EN 60068-2-27 and
EN 60721-3-2
According to EN 24180-2-6 and EN 60721-3-2
IEC/EN 60601-1 (Second Edition)
IEC/EN 60601-1 (Third Edition)
CAN/CSA - C 22.2 No. 60601-1-08
GB4943.1 (non-tropical, altitude < 2000 m)
Protection class I
IP20
CE according to MDD 93/42/EEC (Class I)