When the reference trigger occurs, the DAQ device continues to write
samples to the buffer until the buffer contains the desired number of
posttrigger samples. Figure 2-21 shows the final buffer.
Reference Trigger
Pre-Trigger Samples
Figure 2-21. Reference Trigger Final Buffer

Using a Digital Source

To use ai/ReferenceTrigger with a digital source, specify a source and an
edge. The source can be an external signal connected to any PFI or
RTSI <0..6> pin. The source can also be one of several internal signals on
your DAQ device. Refer to Device Routing in MAX in the NI-DAQmx Help
or the LabVIEW 8.x Help for more information.
Also, specify whether the measurement acquisition stops on the rising edge
or falling edge of the ai/ReferenceTrigger signal.
Figure 2-22 shows the timing requirements of the ai/ReferenceTrigger
source.
Rising-Edge
Polarity
Falling-Edge
Polarity
Figure 2-22. ai/ReferenceTrigger Source Timing Requirements
Post-Trigger Samples
Complete Buffer
t
w
t
= 10 ns minimum
w
2-35
Chapter 2
Analog Input
E Series User Manual