Agilent Technologies 16197A Operation And Service Manual page 19

Bottom electrode smd test fixture
Table of Contents

Advertisement

Below is the general procedure for attaching an adapter to the test fixture. (For details, see
the Operation Manual for each adapter.)
Step 1. Rotate the 7-mm connector on the adapter counterclockwise as viewed from above to
completely retract the connecting sleeve.
Step 2. Gently place the test fixture onto the adapter, aligning the mounting holes with the
mounting posts on the adapter, and the 7-mm connector with that on the adapter.
Step 3. Rotate the 7-mm connector on the adapter counterclockwise as viewed from above to
connect it with the connector on the bottom of the test fixture.
NOTE
Turn the 7-mm connector on the adapter using a 3/4-inch torque wrench with 12 lb-inch
torque (Agilent part number: 8710-1766) to firmly connect the test fixture.
Figure 1-8
Installing the Test Fixture
Chapter 1
Connecting the Test Fixture with a Measuring Instrument
Installation Guide
19

Advertisement

Table of Contents
loading

Table of Contents