Step 1. Order your new or restored A1 Bulkhead Assembly from Table 2-3, "Bulkhead
Step 2. Follow the disassembly and reassembly procedures for Bulkhead removal and
Step 3. Check the FET balance using the procedure described in "FET Balance
Step 4. Place the new calibration sticker on the plastic shell of the Power Sensor.
NOTE

Repair

Power Sensor repair consists of replacing either the A1 Bulkhead Assembly, or the
A2 Power Sensor Board Assembly.

A1 Bulkhead Assembly

Repair Strategy

The recommended repair strategy for the A1 Bulkhead Assembly is to completely
replace it. The replacement Bulkhead is calibrated at the factory, and is supplied
with a calibration report and a new calibration sticker for the Power Sensor.

Procedure

Assembly," on page 31.
replacement. See "Disassembly / Reassembly Procedures" on page 49.
Adjustment" on page 45. If you did not disturb the wires, it is likely that no
adjustment is necessary.
Although the recommended Bulkhead strategy is to completely replace it, we are
aware that some customers have both the ability and experience that enables them to
disassemble and repair Bulkheads to a lower level. For this reason, Appendix B
provides exploded views of the various Bulkheads, and tables listing the parts.
Lower level Bulkhead repair can be more economical than replacement, although
this may be offset by the need to invest in sensor calibration equipment and a SWR
test set-up.