Chapter 2
Analog Input

AI Hold Complete Event Signal

External Strobe Signal

Note
External Strobe is used for signal conditioning with SCXI and is not available for
use with NI-DAQmx.
E Series User Manual
AI Hold Complete Event (ai/HoldCompleteEvent) is an output-only signal
that generates a pulse with the leading edge occurring approximately 50 to
100 ns after an A/D conversion begins. The polarity of this output is
software-selectable, but is typically configured so that a low-to-high
leading edge can clock external AI multiplexers indicating when the input
signal has been sampled and can be removed. This signal has a 400 to
500 ns pulse width and is software-enabled. Figure 2-34 shows the timing
for ai/HoldCompleteEvent.
ai/ConvertClock
ai/HoldCompleteEvent
Figure 2-34. ai/HoldCompleteEvent Timing
External Strobe is an output-only signal on the EXT STROBE pin that
generates either a single pulse or a sequence of eight pulses in the
hardware-strobe mode. An external device can use this signal to latch
signals or to trigger events. In the single-pulse mode, software controls the
level of External Strobe. A 10 ms and a 1.2 µs clock are available for
generating a sequence of eight pulses in the hardware-strobe mode.
Figure 2-35 shows the timing for the hardware-strobe mode External
Strobe signal.
V
OH
V
OL
Figure 2-35. External Strobe Timing
t
d
t
= 50 to 100 ns
d
t
= 400 to 500 ns
w
t
t
w
w
t
= 600 ns or 5 µs