AI Pause Trigger Signal

Note
Pause triggers are only sensitive to the level of the source, not the edge.
© National Instruments Corporation
The output is an active high pulse. Figure 2-8 shows the timing behavior of
the PFI 1/AI REF TRIG pin configured as an output.
Figure 2-8. PFI 1/AI REF TRIG Timing Behavior
The PFI 1/AI REF TRIG pin is configured as an input by default.
You can use the AI Pause Trigger (ai/PauseTrigger) signal to pause and
resume a measurement acquisition. This signal is not available as an output.

Using a Digital Source

To use ai/PauseTrigger, specify a source and a polarity. The source can be
an external signal connected to any PFI or RTSI <0..6> pin. The source can
also be one of several other internal signals on your DAQ device. Refer to
Device Routing in MAX in the NI-DAQmx Help or the LabVIEW 8.x Help
for more information.
Also, specify whether the measurement sample is paused when
ai/PauseTrigger is at a logic high or low level.

Using an Analog Source

When you use an analog trigger source, the internal sample clock pauses
when the Analog Comparison Event signal is low and resumes when the
signal goes high (or vice versa). Refer to the
Source
section of Chapter 10, Triggering, for more information on analog
trigger sources.
t
w
t
= 50 to 100 ns
w
2-19
Chapter 2
Analog Input
Triggering with an Analog
E Series User Manual