Performing Open and Short Correction

Performing Open Correction

The open correction procedure is as follows.
1. Push the HIGH side contact pin so that the distance between the HIGH and the LOW
NOTE
Before performing open correction, remove the DUT used for positioning by pulling back
the lever to release the LOW side contact pin.
2. Tighten the contact pin locking screw to secure the HIGH side contact pin.
Figure 2-3
Contact pin position for open correction
3. Perform the open correction as described in the specific instrument's manual.
16
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
contact pins matches the DUT's width (Figure 2-3).
It is recommended that you place the DUT on the measuring stage and precisely
position the HIGH side contact pin to actual DUT width.