OMEGA Model CYD201/CYD208 User's Manual
F0
Sets Sensor Units for Temperature Display.
Input:
F0x
Returned: Nothing
Remarks: Sets sensor units for the temperature display, where
x = C (Celsius), F (Fahrenheit), K (kelvin), or V (volts).
H
Select High Alarm Setpoint Value.
Input:
Hxxx.x
Returned: Nothing
Remarks: Sets high alarm setpoint, where xxx.x = temperature setpoint
in units specified by F0 command. If the instrument is set for
volts, the alarm defaults to kelvin.
Example:
H300[term] sets a high alarm setpoint of 300 degrees.
L
Select Low Alarm Setpoint Value.
Input:
Lxxx.x
Returned: Nothing
Remarks: Sets low alarm setpoint, where xxx.x = temperature setpoint
in units specified by F0 command. If the instrument is set for
volts, the alarm defaults to kelvin.
Example:
L31.2[term] sets a low alarm setpoint of 31.2 degrees.
R
Alarm Reset.
Input:
R
Returned: Nothing
Remarks: Resets the alarm.
Y
Channel Dwell Time (Model CYD208 Only).
Input:
Yab
Returned: Nothing
Remarks: Sets dwell time for a given channel, where a = channel 1 - 8,
and b = the dwell time parameter as follows:
0 = zero seconds, 1 = 5 seconds, 2 = 10 seconds,
3 = 30 seconds, 4 = 60 seconds
Setting a dwell time of 0 skips the specified channel in the
sequence.
Example: Y23[term] sets the dwell time for channel 2 to 30 seconds.
Remote Operation
4-5