Omega Engineering HH501BR User Manual

Omega Engineering HH501BR User Manual

Omega engineering thermometer user manual

Advertisement

Quick Links

OMEGA

Advertisement

Table of Contents
loading

Summary of Contents for Omega Engineering HH501BR

  • Page 1 OMEGA...
  • Page 2 European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
  • Page 3: Safety Information

    This instrument is a 3½ digit, compact-sized portable digital thermometer designed to use external thermocouples as temperature sensor. Temperature indication follows Reference Temperature/Voltage Tables (N.I.S.T. Mono- graph 175 Revised to ITS-90) for thermocouples. SAFETY INFORMATION It is recommended that you read the safety and operation instructions before using the thermometer.
  • Page 4: Specifications

    SPECIFICATIONS GENERAL Display: 3½ digit liquid crystal display (LCD) with a maximum reading of 1999. Polarity: Automatic, positive implied, negative polarity indication. Overrange: -OL is displayed. Zero: Automatic. Reading Rate: one time per second. Low battery indication: the " below the operating level. Measurement rate: 1 times/second.
  • Page 5 ENVIRONMENTAL Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H. ELECTRICAL Temperature Scale: Celsius or Fahrenheit user-selectable. Measurement Range: 0°C to 1767°C, 32°F to 1999°F Resolution: 1°C, 1°F Accuracy: Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.
  • Page 6: Operating Instructions

    OPERATING INSTRUCTIONS Power Switch key turns the thermometer on or off. In the MIN MAX record mode can not power off, must leave MIN MAX record mode then power off. °C °F Selecting the Temperature Scale Readings is displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F).
  • Page 7 The input selection indicates which input is selected for display; T1 thermocouple, T2 thermocouple or the difference between the two thermo- couples (T1-T2), when the thermometer is turned on, when the thermom- eter is turned on, it is set to the input selected that was in use when the thermometer was last turned off.
  • Page 8 TIMER STOP-WATCH Mode (for HH501AR / HH501AS) 7 8 9 Press TIMER STOP-W key to enter stop-watch mode stop-watch is displayed. Press START/STOP key to toggle stop-watch starts and stops (Time goes up to 19 minutes & 59.9 seconds). Press CLEAR key to display 0.1 second, then press again to zero the stop-watch.
  • Page 9 SET mode (Relative value set and Hi/Lo Limits value set) 1. Press SET key to enter Relative valuse SET mode (Press ENTER key can escape relative valuse set mode), REL set mode. _ = = =.= is displayed (the ".=" is displayed when you set up the fourth number). Relative value is entered via overlay numbers, when you want to get negative values push (- 0) key for end of numbers, then press overlay ENTER key, stored the relative value, enter Hi/Lo Limits value set mode.
  • Page 10 Hi/Lo LIMITS mode Press Hi/Lo LIMITS key to enter the Hi/Lo Limits comparative mode, "LIMIT" is displayed. In the this mode the automatic power-off feature is disabled and other funtion key are disabled, when input temperature value exceed Hi. The beeper emits a continuity pulse tone and "Hi" is displayed, and when input temperature value exceed Lo value.
  • Page 11: Operator Maintenance

    OPERATOR MAINTENANCE WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the thermometer before removing the cover. Battery Replacement Power is supplied by a 9 volt "transistor" battery. (NEDA 1604, IEC 6F22). The " " appears on the LCD display when replacement is needed. To replace the battery, remove the two screws from the back of the meter and lift off the battery cover.
  • Page 12: Calibration Procedure

    HH501(A,B)K / HH501(A,B)J / HH501(A,B)T / HH501(A,B)E CALIBRATION PROCEDURE Note: The following calibration procedure should perform only by qualified technicians who have access to the items as following. Equipment: The class of calibrator had better 10 times greater than the measured meter. 1.
  • Page 13 OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.
  • Page 14 Where Do I Find Everything I Need for Process Measurement and Control? TEMPERATURE PRESSURE/STRAIN AND FORCE FLOW/LEVEL pH/CONDUCTIVITY M-3204 OMEGA...Of Course! DATA ACQUISITION HEATERS ENVIRONMENTAL MONITORING AND CONTROL...