5.7

RECALIBRATION

OMEGA calibrates and certifies thermometers to original factory
specifications for a reasonable fee. You can recalibrate the thermometer to
original specifications, but OMEGA will not warrant these calibrations.
Recalibration requires a digital voltmeter (DVM) with 4½ digit resolution or
better; and 25 k
better.
Recalibration involves current source and analog/digital (A/D) converter
calibration. A/D calibration erases all SoftCal™ compensations; perform it
after current source calibration, not before.
5.7.1

Current Source Calibration

1. Allow 30 minute warm-up to achieve rated specifications.
2. Configure 125 k
channel 1.
3. Connect DVM voltage leads across the resistor and adjust the I ADJ
trimpot (located on rear panel) until DVM displays a voltage of 1.2500
volts ±100 microvolts.
4. Remove DVM and resistor.
Model CYD201
NOTE: Do not use a voltmeter for A/D Converter Calibration.
Service


and 125 k
precision resistors with ±0.01% tolerance or

resistor as shown in Figure 5-4. Set Model CYD208 to
Figure 5-4. Calibration Connections
Model CYD208