SF6 Gas Detection Mode Menu

The Imager can operate in radiometric mode or in gas detection mode. The Imager cannot operate in
both modes at the same time. When SF6 Gas Detection Mode is enabled, the radiometric features of
the Imager are disabled and cannot be selected in the menus.
Table
16
lists the options in the SF6 Gas Detection Mode menu.
Option
SF6 Gas Detection Mode:
ON
SF6 Gas Detection Mode:
OFF
Image Capture
Video Capture
High Gain (Tripod)
Low Gain (Handheld)

Gas Detection Conditions

The Imager detects gas leaks based on:
The temperature difference between the gas and the background environment
The wind speed
The amount of clutter in the background scene (for example, clouds)
The stability of the Imager
The proximity of the Imager to the leak
The lower the volume of the gas leak the more challenging it is for the Imager to detect the leak.
36
Note
Table 16. SF6 Gas Detection Mode Menu
Turns on the gas detection mode. See
Turns off the gas detection mode.
Sets up the Imager to capture an image in gas mode when SF6 Gas Detection
Mode: ON is enabled.
Sets up the Imager to capture a video in gas mode when SF6 Gas Detection
Mode: ON is enabled.
Optimizes the sensitivity of the display when the Imager is mounted on a tripod
when SF6 Gas Detection Mode: ON is enabled. See
Conditions.
Optimizes the sensitivity of the display when the Imager is held in a hand when
SF6 Gas Detection Mode: ON is enabled. See
Note
Description
Gas Detection
Conditions.
Gas Detection
Gas Detection
Conditions.