Calibrating a Pressure Transmitter

The following steps explain calibrating a pressure
transmitter.
1.
Connect the Calibrator to the instrument under test
as shown in Figure 21.
2.
Press
l
press
l
3.
Press
U
information@ITM.com
for current (upper display). If required,
again to activate loop power.
(lower display).
Fluke-Direct
Multifunction Process Calibrator
Calibrating a Pressure Transmitter
M
4.
If necessary, press
for SOURCE mode.
5.
Zero the pressure module.
6.
Perform checks at 0 % and 100 % of span and
adjust the transmitter as necessary.
Note
When HART resistor mode is selected, the
250 Ω resistor is turned on both mA channels.