726
Users Manual
Resistor ON/OFF
HART
1.
Press C until SELECT HART ON or OFF
appears on the display.
2.
Use V to toggle ON or OFF.
3.
Press S to save the setting.
When HART mode is selected, the 250 Ω
resistor is turned on both mA channels.
16
information@ITM.com
Note
Fluke-Direct

Getting Started

This section details some basic operations of the
Calibrator.

Voltage to Voltage Test

To perform a voltage-to-voltage test:
1.
Connect the Calibrator's voltage output to its voltage
input as shown in Figure 6.
O
2.
Press
to turn on the Calibrator. Press
select dc voltage (upper display).
M
3.
If necessary, press
for SOURCE mode (lower
display). The Calibrator is still measuring dc voltage,
the active measurements are visible in the upper
display.
4.
Press
to select dc voltage sourcing.
V
5.
Press
and
to select a digit to change. Press
Y
Z
to select 1 V for the output value. Press and hold
to enter 1 V as the 0 % value.
J
6.
Press
X
to increase the output to 5 V. Press and
hold
to enter 5 V as the 100 % value.
G
7.
Press
and
to step between 0 and 100 %
H
I
in 25 % step increments.