Specifications - Omron GentelTemp 510 Instruction Manual

Digital ear thermometer
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
13413 MC-510-E2
29-10-2008
GentleTemp 510
English
Is the temperatures measured in the right ear different from that
measured in the left ear?
Among healthy people, there should be no significant difference in the
measurement results. Differences may be caused by the following reasons:
1) The infrared sensor is not inserted in the same way.
2) Measurement is not conducted by inserting the unit in a stable manner in
the same angle. Try to measure in either ear that can show the highest
measurement value constantly. In accordance with the "10 seconds mode,"
the unit detects the highest temperature. (Refer to paragraph 3.2).
The temperature shown is rather low.
1 The probe cover is dirty.
2 The infrared sensor is dirty.
3 You removed the unit from the ear before the measurement is finished.
4 The ear is cold. The temperature tends to indicate low when you use
an ice bag or an ice pack, or immediately after coming in from the
cold in winter.
5 The thermometer is not inserted deep enough in the ear.
6 The infrared sensor is not directed towards the eardrum.
Use the 10 seconds measurement mode (Refer to paragraph 3.2).
6

SPECIFICATIONS

Name:
GentleTemp 510
Model:
MC-510-E2
Voltage:
3V DC (lithium battery CR2032)
Power consumption:
0.05 W
zur Verfügung gestellt von Gebrauchs.info
13:22
Pagina 13
®
Aktualisierungsstand Gebrauchs.info November 2011
Battery life:
With a new battery approx. 5000 measurements
Sensor:
Thermopile
Temperature indication:
In 3 digits of +°C, every 0.1°F or °C
Precision:
According to NEN-EN 12470- 5 Clinical
(Celsius)
thermometers – Part 5 Performance of infra-red
ear thermometers (with maximum device)
34°C – 42.2°C: ±0.2°C
Based on the measurement of the standard black
body at the room temperature of 25°C (RH 50 %)
Precision:
According to NEN-EN 12470- 5 Clinical
(Fahrenheit)
thermometers – Part 5 Performance of infra-red
ear thermometers (with maximum device)
93.2°F – 108°F: ±0.3°C
Based on the measurement of the standard black
body at the room temperature of 77°F (RH 50 %)
Measurement range:
34°C - 42.2°C (93.2 - 108°F)
Weight:
Approx. 50 g (1.8 oz) (including the battery)
External dimensions:
Approx. 4.6 cm (B) X 9.3 cm (H) X 5.7 cm (D)
Operating environment:
Ambient temperature; 10°C to 40°C (50°F to 104°F)
Relative humidity:
30 - 85%
Storage environment:
Ambient temperature; -20°C-60°C (-4°F- 140°F)
Relative humidity:
10 - 95%
Accessories:
Storage case, Instruction Manual, probe covers
(10 pieces)
A probe cover is already attached to the main unit.
Options:
Probe cover exclusively for GentleTemp 510
series, MC-PROBE TYPE J
®
13

Advertisement

Table of Contents
loading

Table of Contents