  • Page 1 English Français Deutsch Italiano Español Infrared Nederlands Forehead Thermometer Model Gentle Temp 720 Türkçe Instruction Manual IM-MC-720-E-02-07/2014 0186342-2B...
  • Page 2: Table Of Contents

    Contents Thank you for purchasing the OMRON Gentle Temp 720 Infrared Forehead Thermometer. Intended Use: The OMRON Gentle Temp 720 offers safe, comfortable, and quick temperature measurement from the forehead. It is mainly designed for household use. Important Safety Information ........1 4.
  • Page 3: Important Safety Information

    Important Safety Information To assure correct use of the product, basic safety measures should always be followed, including the precautions listed below. Warning: The following situations are potentially hazardous and, if not avoided, could result in death or serious injury. •...
  • Page 4 Important Safety Information • If the forehead is cold, wait until the forehead is warmed up before taking a temperature measurement. The measured result may indicate low when you use an ice bag or an ice pack, or immediately after coming in from the outside in winter. •...
  • Page 5: Overview

    1. Overview Main unit: Display: Probe lens Surface temperature Room temperature Last reading Probe lens cap Display Battery icon ON/MEM button Mute icon Battery START button Temperature mode (°F) compartment Memory icon Temperature mode (°C) Forehead temperature...
  • Page 6: Preparation

    2. Preparation Removing the Insulating Tape When using the thermometer for the first time, pull the insulating tape out of the battery compartment. Back of the unit The unit is turned on and after 1 minute the room temperature appears on the display. Notes: •...
  • Page 7: Setting The Buzzer

    2. Preparation Setting the Buzzer The buzzer is activated by default, but can be deactivated. Note: The buzzer is only available in forehead measurement mode. Press the ON/MEM button to turn on the unit. Press and hold the ON/MEM button for 3 seconds. The “...
  • Page 8: Using The Unit

    3. Using the Unit Taking a Reading Forehead Measurement Mode This mode displays the oral-equivalent value of the measured forehead temperature. Always make sure the probe lens is clean and undamaged, and the forehead is clean with no sweat, cosmetics, scars, etc.
  • Page 9 3. Using the Unit Hold the probe lens 1 to 3 cm away from the center of the forehead. Notes: • Holding the unit too long in the hand may cause a higher ambient temperature reading of the probe. This could make the body temperature measurement lower than usual. •...
  • Page 10: Measurement Result

    3. Using the Unit Move the unit away from the forehead and check the measurement result. The display lights up for 5 seconds. Notes: • If the result is higher than 42.2°C (108.0°F), “Hi” appears on the display. If the result is lower than 34.0°C (93.2°F), “Lo” appears on the display. •...
  • Page 11: Surface Measurement Mode

    3. Using the Unit Surface Measurement Mode Surface measurement mode shows the actual and unadjusted surface temperature of an object, which is different from the body temperature. This mode can help you determine if the object temperature is suitable for a patient or baby (such as baby milk).
  • Page 12: Using The Memory Function

    3. Using the Unit Using the Memory Function This unit automatically stores the last 25 measurement results (excluding Hi/Lo results). Notes: • If the memory is full, the unit will delete the oldest reading. • When the unit is switched between °C and °F, all the readings stored in the memory are deleted. Press the ON/MEM button to turn on the unit.
  • Page 13: Troubleshooting And Maintenance

    Remove the battery, wait for 1 minute, reinsert the battery, and turn on the unit again. If the message Errors 5–9 indicate a malfunction of the unit. reappears, please contact the OMRON retail outlet or distributor for having the unit checked.
  • Page 14 • Forehead measurement mode: The temperature Use the unit within its measurable range. If a taken is higher than +42.2°C (108°F). malfunction still exists, please contact an OMRON • Surface measurement mode: The temperature retailer. taken is higher than +80°C (176°F).
  • Page 15: Questions And Answers

    4. Troubleshooting and Maintenance Questions and Answers 1. The temperature indicated is rather high. • The unit has been stored in a cold room. When the unit has been stored in a cold room, the measured temperature may be rather high if you measure the temperature shortly after bringing the unit to a room that is at normal temperature.
  • Page 16: Maintenance

    Maintenance • Check the unit if it is dropped. If unsure, please contact the OMRON retail outlet or distributor for having the unit checked. • Clean the probe lens carefully to avoid damaging it.
  • Page 17: Replacing The Battery

    4. Troubleshooting and Maintenance Replacing the Battery Battery: CR2032 lithium button battery Note: To protect the environment, discard the used batteries in accordance with the local regulations regarding waste disposal procedure. Disposal can be done at your retail store or at appropriate collection sites. Release the screw and remove the battery cover.
  • Page 18: Technical Data

    5. Technical Data Product Description: Infrared Forehead Thermometer Model: OMRON Gentle Temp 720 (MC-720-E) Sensor: Thermopile Temperature Display: 4-digit, °F display in 0.1 degree increments 3-digit, °C display in 0.1 degree increments Accuracy: Forehead measurement mode ±0.2°C (±0.4°F) within 35.0°C to 42.0°C (95.0°F to 107.6°F), ±0.3°C (±0.5°F) for other range...
  • Page 19 (such as dropping or physical misuse) caused by the user. Claimed products will only be replaced when returned together with the original invoice/cash ticket. Do not disassemble or modify the unit. Changes or modifications not approved by OMRON HEALTHCARE will void the user warranty.
  • Page 20 Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case the distance is shorter. Further documentation in accordance with EN60601-1-2:2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual. Documentation is also available at www.omron-healthcare.com.
  • Page 21: Some Useful Information

    6. Some Useful Information How the Gentle Temp 720 Works This unit measures the temperature at the surface of the forehead by sensing infrared emission of the skin at the measured point, and then calculates the oral-equivalent value using a unique algorithm based on actual body temperature survey data.
  • Page 22 Forehead temperature (°C) Surface measurement mode Displays the actual temperature at the surface of the forehead. (Survey conducted by Omron) (room temperature at 23°C) 36.5 37 36.5 37.5 38 37.5 38 38.5 39...
  • Page 23 (Cases surveyed: 190, ages: 0–68, environment temperature: 23±2°C) Equivalent Forehead higher Oral higher Number cases (Survey conducted by Omron) 1.4 1.2 1.4 1.2 0.8 0.6 0.4 0.2 0.8 0.6 0.4 0.2 0 0.2 0.4 0.6 0.8 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6...
  • Page 24 Gottlieb-Daimler-Strasse 10, 68165 Mannheim, GERMANY www.omron-healthcare.de OMRON SANTÉ FRANCE SAS 14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE Uniquement pour le marché français: OMRON Service Après Vente Nº Vert 0 800 91 43 14 www.omron-healthcare.fr Made in China Fabricado en China Fabriqué...