Programmable Outputs - Hach BioTector B3500s User Manual

Table of Contents

Advertisement

8.3.5.2

Programmable Outputs

PWR BRD OUT 1-6
PWR BRD OUT 2
DEFAULT STATE
STOP
FAULT
WARNING
NOTE
SAMPLER FILL
...........................
...........................
N/E
*
*
*
*
..
..
▪ POWER BOARD OUTPUT1-6 menus typically contain output
settings related to external devices. These relay outputs, located
on the Power and Input/Output Board, are system optional
features.
▪ POWER BOARD OUTPUT1-6 relays can be programmed to a
single output function or to multiple output functions as
described for OUTPUT 1-6 above. The programmed output
function or functions are marked with an asterisk "*" sign in these
menus as in the example for POWER BOARD OUTPUT 2,
which is programmed for STOP, FAULT, WARNING and NOTE
signals below.
▪ In the example below, POWER BOARD OUTPUT 2 is activated
when STOP or FAULT or WARNING or NOTE signal is
triggered.
▪ POWER BOARD OUTPUT 1 is a standard system relay
programmed to STOP, FAULT, WARNING and NOTE output
functions by default.
▪ In BioTector configuration data download (or in All Data
download), the programmed output functions for each output is
tabulated and marked with asterisk "*" signs for clarity as
illustrated below.