8.3.5.2
Programmable Outputs
PWR BRD OUT 1-6
PWR BRD OUT 2
DEFAULT STATE
STOP
FAULT
WARNING
NOTE
SAMPLER FILL
...........................
...........................
..........................
N/E
*
*
*
*
..
..
..
POWER BOARD OUTPUT1-6 menus typically contain output
settings related to external devices. These relay outputs, located
on the Power and Input/Output Board, are system optional
features.
POWER BOARD OUTPUT1-6 relays can be programmed to a
single output function or to multiple output functions as
described for OUTPUT 1-6 above. The programmed output
function or functions are marked with an asterisk "*" sign in these
menus as in the example for POWER BOARD OUTPUT 2,
which is programmed for STOP, FAULT, WARNING and NOTE
signals below.
In the example below, POWER BOARD OUTPUT 2 is activated
when STOP or FAULT or WARNING or NOTE signal is
triggered.
POWER BOARD OUTPUT 1 is a standard system relay
programmed to STOP and FAULT output functions by default.
In BioTector configuration data download (or in All Data
download), the programmed output functions for each output is
tabulated and marked with asterisk "*" signs for clarity.
Page 125