General Instructions On Charging; How To Charge The Batteries - Invacare TDXSP2X-CG User Manual

Power wheelchair base
Table of Contents

Advertisement

Invacare® TDX® SP2 Series

7.3.2 General Instructions on Charging

Follow the instructions listed below to ensure safe use
and longevity of the batteries:
Charge 18 hours prior to initial usage.
l
We recommend charging the batteries
l
daily after every discharge even after
partly discharge, as well as overnight.
Depending on the level of discharge, it
can take up to 12 hours until the
batteries are fully charged again.
When the battery indicator reaches the red
l
LED range, charge the batteries for at least 16
hours, even if the display reads that the charge
is complete.
Try to provide a 24 hour charge once a
l
week to make sure that both batteries
are fully charged.
Do not cycle your batteries at a low state of
l
charge without regularly recharging them fully.
Do not charge your batteries under extreme
l
temperatures. High temperatures above 86°F
(30°C) are not recommended for charging as
well as low temperatures below 50°F (10°C).
Use only charging devices in Class 2. This class
l
of chargers may be left unattended during
charging. All charging devices which are
supplied by Invacare comply with these
requirements.
You cannot overcharge the batteries when
l
using the charger supplied with your
mobility device, or a charger that has
been approved by Invacare.
Protect your charger from sources of heat such
l
as heaters and direct sunlight. If the battery
charger overheats, charging current will be
reduced and the charging process delayed.

7.3.3 How to Charge the Batteries

Refer to the user manuals for your remote and battery
charger for the position of the charging socket and
further information about charging the batteries.
WARNING!
Risk of Injury or Damage
Using the wrong battery charger may cause
explosion and destruction of batteries. To
avoid injury or damage:
-Only use the battery charger supplied with
your mobility device, or a charger that has
been approved by Invacare.
27
WARNING!
Risk of Injury, Damage or Death
Improper charging may cause injury or
damage.
-Charge battery in a cool, dry and well-
ventilated area.
-DO NOT attempt to charge the battery
by attaching cables directly to the
battery terminals.
-DO NOT attempt to charge the battery and
operate the wheelchair at the same time.
-DO NOT attempt to charge the battery when
the wheelchair has been exposed to any type
of moisture.
-DO NOT sit in the wheelchair while charging
the battery.
-DO NOT leave the charger unattended
when the breaker has tripped. Unplug
and discontinue use immediately. Contact
a qualified service technician or Invacare
for service.
WARNING!
Risk of Injury or Damage
Explosive gases can be generated while
charging. To avoid flammable gas buildup and
injury or damage due to explosion:
-During charging, keep the wheelchair and
battery charger away from sources of
ignition, such as flames and sparks.
-Charge the wheelchair in a space at least
twice the volume of the wheelchair.
WARNING!
Risk of short circuit and electric shock if the
battery charger has been damaged
-Do not use the battery charger if it has been
dropped or damaged.
WARNING!
Risk of Injury, Damage or Death
Improper routing of cord(s) may cause
tripping, entanglement or strangulation hazard
that may result in injury, damage or death.
-Ensure all cord(s) are routed and positioned
properly while charging.
-Avoid positioning cord(s) across areas of high
foot traffic (i.e. aisles, doorways, hallways,
etc.) while charging.
-Close supervision and attention is needed
when operating the wheelchair near children,
pets or people with physical/mental
disabilities.