Configuration of FLOWDATA Response
1st byte
2nd byte
A
B
C
D
8
1
1
2
Details of header division
Data
Name
A
Unused
B
FLOWDATA overflow bit
C
Decimal information
D
TASK No.
E
Unused
F
Unit states (Input)
G
FLOWDATA stop bit
H
TASK judgment result bit
I
Disabled
J
K
L
Unit states (Output)
Data
Measurement data
*1 TASK No.=bit data+1
Example: bit data=10B=2 (dec)
indicates TASK3 data.
*2 The input lines are, from the lower bit, assigned to:
In case of ZS
• Input terminal 4 (Unused. Always set to 0.)
• Input terminal 3
• Input terminal 2
• Input terminal 1
• Input terminal 0 (5 inputs in total.)
1 packet of FLAWDATA
Header division
3rd byte
4th byte
E
F
G
H
I
4
5
1
2
1
Reserved bit
Normally set to 0.Set to 1 when there is an overflow in
FLOWDATA accumulation.
Normally set to 0.0: nm order, 1: mm order
TASK1 to TASK4 (*1)
Always 0
Indicates ON/OFF states of various input lines. (*2)
• All data are set to 1 if FLOWDATA does not flow over subse-
quently. (Always set to 1 for one-shot requests)
• All data are set to 0 if FLOWDATA continues to flow over.
(Until FLOWDATA is stopped after requested (in succession)
Indicates the TASK judgment result of data source. (*3)
Reserved bit
Indicates output states (states of measurement results). (*4)
Signed 32 bit data
When the decimal info division is 0, the unit of the data is nm.
When the decimal info division is 1, the unit of the data is mm
FLOWDATA Reference
Data
division
Meaning of
J
K
L
Data
data
1
1
5
32
Description
Non-procedural Communication Command Reference
Number of
=Total
bits
64 bits
ZS-HL-N Series