Invacare POC1-100C-EU User Manual page 68

Table of Contents

Advertisement

Invacare® Platinum® Mobile
LCD DISPLAY TEXT:
Shut Down
Battery Cold
INDICATORS:
Two audible beeps every 16 seconds
YELLOW alarm indicator illuminated
NOTES:
To clear the alarm, turn the concentrator off. Complete solutions 1, 2, 3, or 4 before restarting.
LCD DISPLAY TEXT:
Shut Down
Oxygen Fault
INDICATORS:
Two audible beeps every 16 seconds
YELLOW alarm indicator illuminated
NOTES:
To clear the alarm, turn the concentrator off. Complete Solution 1 and/or 2 before restarting. The alarm limit has a
tolerance of +/- 2%
Alarm signal generation may be delayed by up to 128 seconds due to the monitoring algorithm used to prevent nuisance alarms.
Call your provider and report the problem if alarm continues.
68
DESCRIPTION:
Battery is too cold to allow the battery
to continue operating.
Concentrator function has shut down.
DESCRIPTION:
The oxygen output purity has fallen to
a value below 73%.
Concentrator function has shut down.
SOLUTIONS:
1. Move concentrator to warmer
surroundings. Allow concentrator to
warm up to 41°F (5°C), OR
2. Use AC or DC power and remove
battery, OR
3. Insert an alternate charged battery
and remove the cold battery, OR
4. Change to another source of oxygen
if no other power source is available.
SOLUTIONS:
1. Verify concentrator is in
recommended environmental
temperature. Refer to Specifications
in Technical Data.
2. Clean intake filter and ensure both
intake and exhaust are not blocked.
3. Change to another source of oxygen
if alarm continues.
1187374-F