7.2.2 Shut Down Alarms

The concentrator shuts down when the alarm conditions in this section occur.
LCD DISPLAY TEXT:
Shut Down
No Breath Detected
NOTES:
This alarm does not apply during warm up.
To clear alarm, turn the concentrator off. Complete solution 1 before restarting. Press and hold the On/Off button to power
down the concentrator and restart.
LCD DISPLAY TEXT:
Shut Down
Stuck button
INDICATORS:
Two audible beeps every 16 seconds
YELLOW alarm indicator illuminated
NOTES:
To clear alarm, turn the concentrator off. Complete solution 1 before restarting. Call you provider and report the problem
if alarm continues.
1187374-F
DESCRIPTION:
Oxygen concentrator has not detected a
breath for over a two minute period.
DESCRIPTION:
A stuck button is detected upon
connection to power source.
SOLUTIONS:
1. Verify cannula is connected, not
kinked, properly positioned and user
is breathing through his/her nose.
2. Change to another source of oxygen
if alarm continues.
SOLUTIONS:
1. Remove all power sources from
concentrator. Press each control
button looking for a stuck button.
Reconnect power.
2. Change to another source of oxygen
if alarm continues.
Troubleshooting