External Dc Power; Carry Bag Setup - Invacare POC1-100C-EU User Manual

Table of Contents

Advertisement

The external power input connector of the
concentrator is angled slightly upward. The outlet
connector of the power cord needs to be angled
slightly upward when installed.
When a battery is installed, external AC or DC power
can be connected or disconnected whether the
concentrator is off or on.
When an external power source is properly
connected to the concentrator, the orange indicator
light next to the plug symbol on the control panel
will illuminate and there will be an audible beep.
When external power is disconnected, the orange
light will turn off and there will be an audible beep.

3.3.6 External DC Power

The DC power cable G allows the oxygen concentrator to
be connected to an automobile's (boat, motor home, etc.)
12–volt DC outlet. Use of the DC power cable will allow
1187374-F
the oxygen concentrator to be operated and simultaneously
recharge the battery.
Complete the following steps to connect the DC power cable
to the concentrator:
1. Insert the DC power cable outlet connector F into the
external power input connector C of the concentrator.
2. Insert plug H into a DC outlet after the automobile
(boat, motor home, etc.) is running.

3.4 Carry Bag Setup

CAUTION!
Risk of Damage
– Improper installation of the concentrator into
the carry bag can cause a high temperature
alarm condition and result in the concentrator
shutting down.
– The carry bag is not designed to protect the
concentrator from excessive drops, impacts,
or abuse.
– To install or remove the carry bag the
concentrator must be off and disconnected
from external power and cannula.
Concentrator can be operated with or without
the carry bag. When out of the carry bag, the
concentrator is limited to stationary use. Refer to
Using the Carry Bag in Usage.
Setup
31