Tools and auxiliary devices required • Standard installation tools • Water level • M6 washer (hole diameter 6 - 6.5 mm) Siemens AG Register RXL2-120.071.01 Page 1 of 2 LITHOSTAR Multiline Medical Engineering Rev. 04 05.97 TD SP 1...
Page 6
1 - 2 General This page intentionally left blank. LITHOSTAR Multiline Register RXL2-120.071.01 Page 2 of 2 Siemens AG Rev. 04 05.97 TD SP 1 Medical Engineering...
The washer must appear centered on the marking (smallest format on the adhesive tape). • If necessary, shift the washer on the I.I. input and repeat the entire procedure, until the "electrical center" is determined. (tolerance ± 1 mm) LITHOSTAR Multiline Register RXL2-120.071.01 Page 2 of 10 Siemens AG Rev. 01 05.97...
Loosen the two screws with a 6mm Allen key, but do not remove them. The rubber panelling can now be removed towards the top. • Reinstall the rubber panelling in the reverse order. Siemens AG Register RXL2-120.071.01 Page 3 of 10 LITHOSTAR Multiline Medical Engineering Rev. 01 05.97 TD SP 1...
Remove the 7 screws (1/Fig. 5) on each side. • The I.I. can be shifted with the screw (2/Fig. 5) in the patient transverse direction. • Retighten the screws (1/Fig. 5). LITHOSTAR Multiline Register RXL2-120.071.01 Page 4 of 10 Siemens AG Rev.
Go into the ”uncontrolled mode” as described in the chapter ”System Logon, Adjustment Calibration” of the operating instructions for service software (RXL2-120.113.01...). Siemens AG Register RXL2-120.071.01 Page 5 of 10 LITHOSTAR Multiline Medical Engineering Rev. 01 05.97 TD SP 1...
Page 12
- the switch must be closed in the range between 4.75V and 4.79V, (ADC value 954 ± 4 ) - the switch must be open in the range between <4.73 V (ADC value > 946) and > 4.81 V (ADC value > 962). LITHOSTAR Multiline Register RXL2-120.071.01 Page 6 of 10 Siemens AG Rev.
• Remove the adjustment device (does not apply to replacement of the I.I.) • Perform the cross hairs adjustment or final steps. Siemens AG Register RXL2-120.071.01 Page 7 of 10 LITHOSTAR Multiline Medical Engineering Rev. 01 05.97 TD SP 1...
Check whether the image orientation is vertical or horizontal on the monitor. • To adjust, loosen the mounting screws on the camera and turn the camera head. • Check the image orientation again with FL . LITHOSTAR Multiline Register RXL2-120.071.01 Page 8 of 10 Siemens AG Rev.
Perform an IQ test according to chapter 6 "Dose rate" of the Start-up Instructions (does not apply when replacing the shock wave head). • Replace or supplement the documents. • Check all functions. Siemens AG Register RXL2-120.071.01 Page 9 of 10 LITHOSTAR Multiline Medical Engineering Rev. 01 05.97 TD SP 1...