The Braun ThermoScan thermometer has been carefully developed for accurate, safe and fast temperature measurements in the ear. The shape of the thermometer prevents it from being inserted too far into the ear canal to hurt the eardrum.
However, as with any thermometer, proper technique is critical to obtaining accurate temperatures. Therefore, read the use instructions carefully and thoroughly.
Braun ThermoScan measures the infrared heat generated by the eardrum and surrounding tissues. To avoid gross temperature differences, the sensor itself is brought to a temperature close to that of the human body. When the Braun ThermoScan is placed in the ear, it continuously monitors the infrared radiation. The measurement will be finished and the result is displayed, when an accurate reading can be assured.
The goal of thermometry is to measure core body temperature which is the temperature of the vital organs. Ear temperatures accurately reflect core body temperature, since the eardrum shares blood supply with the temperature control center in the brain, the hypothalamus. Therefore, changes in body temperature are reflected sooner and more accurately in the ear than at other sites.
Normal body temperature is a range. The following table shows that this normal range also varies by site. Therefore, readings from different sites should not be directly compared.
Normal ranges by site: | ||
Axillary: | 34.7 – 37.3°C | 94.5 – 99.1°F |
Oral: | 35.5 – 37.5°C | 95.9 – 99.5°F |
Rectal: | 36.6 – 38.0°C | 97.9 – 100.4°F |
ThermoScan: | 35.8 – 38.0°C | 96.4 – 100.4°F |
Normal ThermoScan ranges by age: | ||
0 – 2 years | 36.4 – 38.0°C | 97.5 – 100.4°F |
3 – 10 years | 36.1 – 37.8°C | 97.0 – 100.0°F |
11 – 65 years | 35.9 – 37.6°C | 96.6 – 99.7°F |
> 65 years | 35.8 – 37.5°C | 96.4 – 99.5°F |
However, the range of normal also varies from person to person and fluctuates throughout the day. It is therefore important to determine your normal temperature range. This is easily done using Braun ThermoScan. Practice taking temperatures on yourself and healthy family members to determine their normal temperature ranges.
Note: When consulting your physician, communicate that the ThermoScan temperature is a temperature measured in the ear and if possible, note the individual's normal ThermoScan temperature range as additional reference.
The last temperature taken is stored in its memory and will be automatically displayed when it is turned on again. The display will show «MEM».
IRT 4520:
This model stores the last 8 temperature readings. To display the stored readings, the thermometer must be turned on. Then press the «I/O» button for at least 1 second. The display shows the memory number (e. g. MEM 1), and when releasing the «I/O» button, the stored temperature for that memory number is displayed, together with «MEM». If «I/O» button is pressed too long, the thermometer will be switched off. Each further pressing of the «I/O» button displays the remaining memory numbers (up to MEM 8).
MEM 1 is the latest reading, MEM 8 would be the oldest.
The memory mode is left automatically after displaying the oldest reading, or after pressing the «I/O» button for at least 1 second.
Your Braun ThermoScan is shipped with the Celsius (°C) temperature scale activated. If you whish to switch to Fahrenheit (°F) and/or back from Fahrenheit to Celsius, proceed as follows:
The probe tip is the most delicate part of the thermometer.
It has to be clean and intact to ensure accurate readings.
If the thermometer is ever accidentally used without a lens filter, clean the probe tip as follows: Very gently wipe the surface with a cotton swab or soft cloth moistened with alcohol. After the alcohol has completely dried out, you can put a new lens filter on and measure.
If the probe tip is damaged, contact Kaz.
Use a soft, dry cloth to clean the thermometer display and exterior.
Do not use abrasive cleaners. Never submerge this thermometer in water or any other liquid.
Store thermometer and lens filters in a dry location free from dust and contamination and away from direct sunlight.
Additional lens filters (LF 40) are available at most stores carrying Braun ThermoScan or at Kaz Service Centers.
Error message | Situation | Solution |
No lens filter is attached. | Attach new, clean lens filter. | |
The thermometer cannot identify a secure position of the probe. An accurate measurement was not possible. POS = position error | IRT 4020: Clear the display by pushing the «start» button once. IRT 4520: Clear the display by pushing the «I/O» button once. Take care that the positioning of the probe is correct and remains stable. | |
Ambient temperature is not within the allowed operating range (10 – 40°C or 50 – 104°F). | Allow the thermometer to remain in a room for 30 minutes where the temperature is between 10 and 40°C or 50 and 104°F. | |
Temperature taken is not within typical human temperature range (34 – 42.2°C or 93.2–108°F). HI = too high LO = too low | Make sure a new, clean lens filter is attached and thermometer is properly inserted. Then, take a new temperature. | |
System error – self-check display flashes continuously and will not be followed by the ready beep and the ready symbol. | Wait 1 minute until the thermometer turns off automatically, then turn on again. | |
If error persists, | ... reset the thermometer by removing the batteries and putting them back in. | |
If error still persists, | ... have the thermometer checked at a Kaz Customer Service Center. | |
Battery is low, but thermometer will still operate correctly. | Insert new batteries. | |
Battery is too low to take correct temperature readings. | Insert new batteries. | |
Do you have any further questions? | Please contact an authorized Service Centre (See Package Insert). |
The thermometer is supplied with two 1.5 V type AA (LR 06). For best performance, we recommend Duracell® alkaline batteries. Insert new batteries when the battery symbol appears on the display.
Open the battery compartment. Remove the batteries and replace with new batteries, making sure the poles are in the right direction. Slide battery door back until it snaps in place.
To protect the environment, dispose of empty batteries at your retail store or at appropriate collection sites according to national or local regulations.
The thermometer is initially calibrated at the time of manufacture. If this thermometer is used according to the use instructions, periodic re-adjustment is not required. If at any time you question the accuracy of the temperature measurement, please contact an authorized Kaz Service Center.
Manufacturing date is given by the LOT number located in the battery compartment. The first number after LOT represents the last digit of the year of manufacture. The next three digits give the day of the year of manufacture. An example: LOT 0116 xx xx – this product has been manufactured on the 116th day of the year 2000.
Displayed temperature range: | 34–42.2°C (93.2–108°F) |
Operating ambient temperature range: | 10–40°C (50–104°F) |
Display resolution: | 0.1°C or °F |
Accuracy for displayed temperature range: | ± 0.2°C (35.5–42°C / 95.9–107.6°F) ± 0.3°C (outside this temperature range) |
Clinical repeatability: | ± 0.14°C (± 0.26°F) |
Battery life: | 2 years / 1000 measurements |
Type BF equipment
Subject to change without notice.
This appliance conforms to the following standards:
DIN EN 60601-1: 3/96 «Medical electrical equipment» –
Part 1: General requirements for safety
DIN EN 12470-5: 2003 «Clinical thermometers« –
Part 5: Performance of infrared ear thermometers (with maximum device)
This product conforms to the provisions of the EC directive 93/42/EEC (Medical Device Directive)
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For detailed description of EMC requirements please contact an authorized local Service Centre (See Package Insert).
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
Please do not dispose of the product in the household waste at the end of its useful life. Disposal can take place at your local retailer or at appropriate collection points provided in your country.
Guarantee
Kaz grants 2 years guarantee on the product commencing on the date of purchase. Within the guarantee period Kaz will eliminate, free of charge, any defects in the appliance resulting from faults in materials or workmanship, either by repairing or replacing the complete appliance as we may choose.
This guarantee extends to every country where this appliance is supplied by Kaz or its appointed distributor.
This guarantee does not cover: damage due to improper use, normal wear or use as well as defects that have a negligible effect on the value or operation of the appliance. The guarantee becomes void if repairs are undertaken by unauthorised persons and if original Braun parts are not used.
To obtain service within the guarantee period, hand in or send the complete appliance with your sales receipt to an authorised Kaz Service Centre.
For UK only:
This guarantee in no way affects your rights under statutory law.
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.