Konica Minolta mKDR Xpress Operation page 68

Digital mobile x-ray unit
Table of Contents

Advertisement

Mobile X-ray Unit
Operation
60
" Battery Critical Low:
This alert appears when the system batteries are at Critically Low Charge
status.
The lower indicator flashes orange, on one or both columns, and the middle and
upper indicators are off, with the unit disconnected from mains (not charging).
This alert indicates that it is necessary to connect the system to a power source
for battery charging, although the operation is limited:
If the indicator is on the X-ray column, indicates that exposures are
disabled.
If the indicator on the Motion column, indicates that motion is enabled for
30 minutes or 1 km after which, all the indicators will be OFF.
The system initiates a shutdown with a 30 minutes countdown, that can
be stopped connecting the unit to the mains.
After 25 minutes without connection, the Status Light Indicator on the
Head-Assembly flashes orange.
If the system is still unplugged when the countdown expires, the unit will
automatically shut down.
OM-0481R1