Mobile X-ray Unit
Operation
ID
Remote Handswitch is out of the placement longer
290402
than configuration time
290454
Internal error located in the head
usability_setting.xml is outdated. Please, enter in
290500
service mode and update this file
positioner_1.xml is outdated. Please, enter in
290501
service mode and update this file
Inhibit X-ray due to current RFID tag does not allow
290700
X-ray
290701
Inhibit X-ray due to screen has been powered off
290702
Inhibit X-ray due to parking position
290703
Inhibit X-ray due to system is moving
290704
Inhibit X-ray due to brakes are released
290705
Inhibit X-ray due to board configuration pending
290706
Inhbit X-ray due to System in Critical Battery Level
290707
Inhibit X-ray due to Dosimeter is not ready
290708
Inhibit X-ray due to Collimator filters are moving
290709
Inhibit X-ray due to tube thermostat is released
290710
Inhibit X-ray due to tube fan is released
290711
Fan Tube released
Fine positioning buttons pressed longer than
290712
allowed
120
DESCRIPTION
TYPE
Return the remote Handswitch to its placement or click
Information
on "Accept button" to disable the warning.
Error
Reboot the system and try to repeat the operation.
Warning
- -
Warning
- -
Inhibit
Exposure
Exposure
Inhibit
Check if unit is in parking position. If it is not, call
Exposure
Service.
Inhibit
Inhibit
- -
Exposure
Inhibit
Call Service.
Exposure
Inhibit
If after charging the unit the error persists, call Service
Exposure
Inhibit
Call Service.
Exposure
Inhibit
- -
Exposure
Wait for the X- -ray Tube to cool.
Inhibit
Exposure
If the error persists, call Service
Inhibit
Call Service.
Exposure
Information
No user action required.
Information
Release the Fine positioning buttons.