Start Of A Measurement Series - Hach GANIMEDE P User Manual

Laboratory machine
Table of Contents

Advertisement

Operation
4.2.1
Menu structure
Options
About
Reagent status
Reagent status
Reagent status 100
Date/time
Display contrast
Lighter
Darker
Interfaces
PC/scanner
Scanner
XON / XOFF
9600
None
LIMS
9600
None
Printers
Language
4.2.2

Start of a measurement series

42
Display of the program version, control files and operating hours
Display of the reagent quantity and the range
Display of the number of samples that can be analyzed with the remaining reagent
quantity.
Reset of the sample counter after the analysis unit has been loaded with a new
reagent kit.
Input of date and time for the open-loop control unit.
Set the display contrast.
Configuration of the interfaces for PC/scanner and LIMS system
Baud rate
Parity
Baud rate
Parity
Display of the connected printers on the open-loop control unit and analysis unit
Selection list of available languages
1. Load the sample tray according to the sample batch with the samples starting from
position 1.
Note: The data is overwritten if a sample batch is already stored. A copy must be created
beforehand if the data is to be retained.
2. If a multi-point calibration is selected for the calibration of the measurement series,
the device refers to one of the three saved calibration curves.
For Ganimede P, the following applies:
The standards are placed on the sample tray with the measurement series as follows
for a two-point calibration:
Position 35: zero solution, GCZ 100
Position 36: standard, e.g. 0.2 mg/L PO
Alternatively:
Position 35: 1st standard, e.g. 0.2 mg/L PO
Position 36: 2nd standard, e.g. 2 mg/L PO
For Ganimede N, the following applies:
The standards are placed on the sample tray with the measurement series as follows
for a two-point calibration:
Position 33: 1st low standard, e.g. 0.5 mg/L TN
Position 34: 2nd low standard, e.g. 30 mg/L TN
Position 35: 1st high standard, e.g. 30 mg/L TN
-P GCS 050
4
-P GCS 050
4
-P GCS 100
4
GCZ 100
B
GCS 050
B
GCS 050
B

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ganimede n

Table of Contents