26
6. Install the suction tube for the rinse water of the sample needle.
To do so, connect the end with a screw thread to the permanently mounted tube on
the rear side of the sample changer.
Insert the free end of the suction tube into a bottle of distilled water.
Note: It is essential that the needle rinser is connected in order to make sure that a sample
batch drains faultlessly. Sufficient rinse water (distilled water) must be available at all times.
7. Install the drain hose.
Attach one end of the suction tube to the white tube connection on the rear side of
the sample changer.
Place the free end of the drain hose in a suitable waste container.
Note: The waste container and the drain hose must always be positioned lower than the sample
changer. The rinsing fluid must be able to drain unimpeded; the drain hose must not be kinked!
Check the capacity of the waste bottle.