Figure 3. Insulation and Substitute Leakage Test Connections Class II
2. Push to start the test:
•
Single test
•
Continuous test
3. The display shows the test progress.
Continuous test only:
4. Push to stop the test.
5. Push to store the test result, if required.
6. For Class II, continue the test for all exposed metal parts on the appliance.
Substitute Leakage Current Test (I
The test measures the leakage current between
•
the earth pin of the appliance cable plug (Class I)
or
•
the test probe to be applied to the appliance under test (Class II) and the Live and Neutral
pins of the appliance (pins are connected together within the Tester for this test).
Refer to the standards and guidance material for the safe applicability of this test.
To perform the Substitute Leakage Current test:
1. Connect the appliance and the probe as shown on the Tester. See Figure 3.
•
For Class I appliances no probe is required.
•
For Class II appliances apply the probe to any exposed metalwork on the appliance.
push momentarily
hold down >2 seconds
)
SUBSTITUTE
Appliance Tester
Safe Appliance Tests
hdc06.eps
13