Drive The External Fan - Agilent Technologies TwisTorr Medium-TMP Rack User Manual

Table of Contents

Advertisement

Technical Information

Drive the External Fan

There are three different ways to drive the external Fan selectable via Front Panel
and via RS232/RS485:
1
Always ON
Setting WIN 143 = 0 (default value) the fan remains always ON (the value in
WIN144 is not considered).
2
Automatic Setting
WIN 143 = 1 the external fan is always ON during the Ramp-up status. In the
others status, the fan behavior depends only on WIN 144 value: WIN 144 = 1
external fan ON; WIN 144 = 0 external fan OFF that means that WIN 144
determines the Fan status when the pump reaches the "normal" status (Front
panel display HICOMP or HIFLOW).
3
Serial setting
WIN 143=2 the external fan is configured in serial mode, in this condition the
fan behavior depends only on WIN 144 value) WIN 144 = 1 external fan ON; WIN
144 = 0 external fan OFF. The WIN 143 default value is 0.
Agilent TwisTorr Medium-TMP Rack Controller User Manual