Rs 232/Rs 485 Communication Description; Communication Format; Communication Protocol - Agilent Technologies TwisTorr Medium-TMP Rack User Manual

Table of Contents

Advertisement

Technical Information

RS 232/RS 485 Communication Description

Both the RS 232 and the RS 485 interfaces are available on the connector P2.
The communication protocol is the same (see the structure below), but only the RS
485 manages the address field. Therefore, to enable the RS 485 is necessary to
select the type of communication as well as the device address by means of the
serial communication software.

Communication Format

8 data bit
no parity
1 stop bit
baud rate: 600 / 1200 / 2400 / 4800 / 9600 / 19200 / 38400 programmable
(default setting: 9600).

Communication Protocol

The communication protocol is a MASTER/SLAVE type where:
Host = MASTER
Controller = SLAVE
The communication is performed in the following way:
the host (MASTER) send a message + CRC to the controller (SLAVE);
the controller answer with an answer + CRC to the host.
The MESSAGE is a string with the following format:
<STX>+<ADDR>+<WIN>+<COM>+<DATA>+<ETX>+<CRC>
Where:
When a data is indicated between two quotes ('...') it means that the indicated data
NOTE
is the corresponding ASCII character.
168
Agilent TwisTorr Medium-TMP Rack Controller User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

X3501-64006X3501-64016

Table of Contents