Instructions for Use

Set-up

For user safety, the Controller must be powered by a 3-wire power cord
WARNING
(see Accessories and Spare Parts Table) and plug (internationally
approved). Always use this power cord and insert the plug in a socket with
a suitable ground connection to avoid electrical shock and to comply with
EC requirements . The high voltage developed inside the controller can
cause severe injury or death. Before installing or servicing the unit, always
disconnect it from the socket.
Power supply cord: the correct cable for electrical wiring is a three wires
WARNING
(Ph+N+Earth) cable.
The wire section has to be at least 1.0 mm
voltage or 2 mm
length equal to 3 m.
Do not install or use the controller in an environment exposed to atmospheric
NOTE
agents (rain, snow, ice), dust, aggressive gases, or in explosive environments or
those with a high fire risk.
During operation, the following environmental conditions must be complied with:
temperature: from 0 °C to +40 °C
relative humidity: 0 – 95 % (non-condensing).
In any case it is necessary that the cooling air can circulate freely around the
apparatus.
136
2
in case of 120 Vac supply voltage for a reference cable
Agilent TwisTorr Medium-TMP Rack Controller User Manual
2
in case of 240 Vac supply