Instructions for Use
The TwisTorr Medium-TMP Rack Controller belongs to the second installation
CAUTION
(or overvoltage) category of the EN 61010-1 standard. Therefore, connect the
device to a mains line that complies with the above category. The TwisTorr
Medium-TMP Rack has Input/Output and serial communication connectors that
must be connected to external circuits in such a way that no live parts of
electronic board are accessible. Make sure that the insulation of the device
connected to the TwisTorr Medium-TMP Rack is adequate even in the case of a
single fault as per the EN 61010-1 standard.
For installation of optional accessories, see "Technical Information" section.

TwisTorr Medium-TMP Rack Controller Mounting

This paragraph details the main operating procedures. Prior to operating the
Controller, make all vacuum manifold and electrical connections, referring to the
Turbo pump instruction manual.
To avoid damage to people and damage to the equipment, before starting
WARNING
the pump, make sure that you have correctly fastened the pump to the
vacuum system, according to the instructions given in the pump manual
(chapter Installation).
Once connected to the line voltage, to operate the controller (and then to start the
NOTE
pump) in a very simple way, connect on the J1 connector (Input-Output)
the pre-wired mating connector that is supplied together with the controller.
The pre-wired mating connector is a simple and quick way to allow the controller
to be operated without having the need to implement other wired external
connections like Serial Com or onboard I/O.
The forepump and turbo pump can be switched on at the same time.
Agilent TwisTorr Medium-TMP Rack Controller User Manual
137