4
Physiological imaging
Applying ECG electrodes
Performing the examination
Evaluating the signal
98
5
Then dry the skin with a paper towel.
6
Use a suitable gel to prepare the skin for better signal transmission.
Finding the best position for the electrodes is a matter of trial and
error.
(
Page 95 Description)
Patients with an offset heart axis (e.g. dilatative cardiomyopathy)
may require a different orientation than parallel to the spine.
1
Check the expiration date of disposable electrodes and order new
ones if necessary.
2
Pull the protective foil off the electrodes and attach them.
3
Use the application cushion and position it together with the PERU
on the patient.
4
Connect the electrode clips of PERU to the ECG electrodes.
Perform the examination. See:
The signal intensity is indicated on the table display for each ECG
channel by stars (max. 5 stars).
For example, a maximum signal intensity of ECG channel I and a low
signal intensity of channel II are shown as follows:
I: *****
II: *
1
Check to see if the leads show a preferably pronounced R-wave in
the dialog box Physiological display or in the PMU display.
2
Ensure that at least 2 stars are shown on one ECG channel (l or II)
of the table display. The other ECG channel must show at least 1
star.
3
If you are using the mMR Body, position it over the heart. Connect
the coil and secure it with the belts. See:
4
Wait at least 10 heart beats (learning phase for triggering) before
you move the patient table into the magnet.
(