4
Physiological imaging
Attaching the PERU
94
WARNING
Physiological displays are not approved to monitor vital
parameters!
Anomalies of the vital parameters may not be recognized or
recognized too late
◆
Never use the physiological displays (either at the system or
at the computer) to monitor the vital parameters of a patient.
◆
Only use suitable patient monitoring systems for monitoring
vital parameters (MR Safe or MR Conditional).
◆
Briefly press the respective button (several times, if necessary) until
you reach the requested setting.
The PERU is used together with ECG and respiratory triggering.
✓
ECG electrodes are attached.
✓
The patient table is in the Home position.
CAUTION
Hot ECG cables!
Patient burns
◆
Place absorbent natural material between the ECG cables/
leads of the PERU and the patient's skin.
1
Position the patient on the patient table with the head toward the
magnet bore.
2
Position the PERU in the application cushion.
Especially with respect to whole-body examinations, it should be
noted that artifacts (homogeneity distortions) may occur in the
direct vicinity of the PERU transmitter unit.
3
Use the application cushion and position it together with the PERU
on the patient.
MR System | Operator Manual
Print No. MR-01501G.621.02.02.02