3
MR system components
Rescue with emergency
unlocking mechanism
Reset the emergency release
64
(1) Unlocking handle
(2) Reset button
1
Pull the unlocking handle outward up to the end stop.
A reset button moves out of the unlocking handle.
The Table Stop function is released and the Table Stop button
glows. The MR measurement is terminated and the table
positioning is deleted. The two movement arrows on the table
display blink alternately.
2
Pull the tabletop out of the magnet using the handle at the foot
end.
3
Rescue the patient.
The error regarding the power supply and/or the motorized drive
has been removed.
1
Press the reset button into the unlocking handle.
The unlocking handle snaps back into its original position.
2
Shift the tabletop manually by approx. 10 cm in any direction until
it audibly locks into the motorized drive unit.
3
Press the Table Up/Inward button.
4
Press the Table Down/Outward button.
The Table Stop is cancelled.
5
Press the Home Position button on the control unit.
The tabletop drives at reduced speed into the Home Position. After
reaching the Home Position, the patient table is ready for operation
again.
MR System | Operator Manual
Print No. MR-01501G.621.02.02.02