5
MR system operation
120
The dialog box offers examination parameters that you can adapt to
allow the examination to continue. You can also change the
operating mode using the button provided.
In the First Level Controlled Operating Mode, often only minor
modifications (if any) may be required.
Online
monitoring: The system constantly measures the transmit
power and ensures that the appropriate limit values are observed.
Examinations in progress will be aborted if the limit is exceeded.
Limit
values: The SAR limits used by the Look Ahead monitoring
function are set according to country-specific approval guidelines at
the time of the syngo MR installation.
Normal operating
barely notices the effects of the RF field. The stress on the cardio-
vascular system is low in general.
mode: In the Normal Operating Mode, the patient
MR System | Operator Manual
Print No. MR-01501G.621.02.02.02