Recalling Readings From Memory; Storing Meter Configuration Information - Fluke 8846A/C User Manual

Table of Contents

Advertisement

8845A/8846A
Users Manual
Each memory file will store up to 10000 readings. If the number of samples
is set to a value greater than 10000, consecutive memory files will be used
to store all the samples. If the readings fill the last file (999), reading
storage will stop.

Recalling Readings from Memory

To recall readings from internal memory:
1. Press M.
2. Press the RECALL READING soft key.
To recall readings from external memory (8846A Only), the memory device must be
removed from the 8846A and plugged into a PC where the comma delimited files can be
read. Each file is labeled MEAS0XXX.CSV. The XXX is the file number starting at 001
and going up to 999. Each file is date and time stamped.

Storing Meter Configuration Information

Up to five meter configurations can be stored in the Meter's internal memory. The
8846A, with optional USB memory installed, can store an additional 99 configurations in
the external memory.
To store a meter configuration in the Meter's internal memory:
1. Press M.
2. Press the STORE CONFIG soft key as shown below.
3. Press the STORE INT MEM soft key.
4. Press one of the five soft keys labeled with the desired memory location to store the
To store a meter configuration in to optional external memory (8846A Only):
1. Press M.
2. Press the STORE CONFIG soft key.
3. Press the STORE USB soft key.
3-20
The Meter displays the first stored reading from the internal file. Four soft keys
provide the means to scroll through the readings stored in the file. The FIRST soft
key displays the first reading in the file while the LAST soft key displays the last
reading. Use the <-- and --> soft keys to move forward and backward through the
file one reading at a time.
+
0.5993
F1
present Meter configuration.
The Meter labels the first four soft keys with the first four memory locations.
CONFIG01 through CONFIG04. The fifth soft key is labeled MORE to allow
access to all 100 memory locations.
Note
F2
F3
F4
F5
caw033.eps

Advertisement

Table of Contents
loading

This manual is also suitable for:

8846a8845a8845a/c

Table of Contents