Current-Input Fuses - Fluke 8846A/C User Manual

Table of Contents

Advertisement

Line Voltage Selection
100
120
220
240

Current-Input Fuses

The 400 mA and 10 A inputs are protected by user-replaceable fuses.
The 400 mA input is protected by a fuse (F2) rated at 440 mA, 1000 V (fast blow),
10,000 A minimum breaking capacity (Fluke Part No. 943121).
The 10 A input is protected by a fuse (F1) rated at 11 A, 1000 V (fast blow),
10,000 A minimum breaking capacity (Fluke Part No. 803293).
For protection against fire or arc flash, replace a blown fuse
only with one from Fluke.
To test for a blown Current Input Fuse:
1. With the Meter powered up, plug a test lead into the VΩGER connector.
2. Press N.
3. Insert the other end of the test lead into the 400 mA input connector.
If the fuse is good, the Meter will read less than 200 Ω. If the fuse is blown, the
Meter will read overload.
4. Remove the probe from the 400 mA connector and insert into the 10 A
connector.
Table 2-1. Line Voltage to Fuse Rating
Fuse Rating
0.25 A, 250 V (slow blow)
0.25 A, 250 V (slow blow)
0.125 A, 250 V (slow blow)
0.125 A, 250 V (slow blow)
Figure 2-1. Replacing the Line Fuse
W Warning
Preparing the Meter for Operation
Power Considerations
Fluke Part No.
166306
166306
166488
166488
caw0201f.eps