To copy the last or interrupted readings file from internal memory to an already-inserted
USB memory device:
1. Press M.
2. Press the INIT MEAS TO USB soft key.
The copy process will take several seconds.
To erase USB memory, refer to the "Media Memory" section in Chapter 1.
To clear internal memory contents:
3. Press M.
4. Press the MANAGE MEMORY soft key as shown below.
5. Press the ERASE MEMORY soft key.
6. If you are sure you want to clear all stored readings, all stored configurations, the
user string and the hostname from internal memory, press the ERASE soft key. If
not, press the CANCEL soft key.
To check the amount of external memory available (8846A Only):
1. Press M.
2. Press the MANAGE MEMORY soft key.
3. Press the USB STATUS soft key.
After a few seconds, the Meter displays the total external memory, the amount of
external memory used and free external memory.

Controlling System-Related Operations

Identifying Meter Errors

When the Meter detects an error, the error annunciator (item 5 in Table 3-2) is
illuminated, and the beeper sounds. A list of possible Meter errors can be found in
Appendix B of this manual.
To read the error or errors:
1. Press I.
2. Press SYSTEM.
3. Press ERROR.
4. The first error, if there is more than one, is displayed. To read additional errors, press
NEXT.
If you want to erase all error messages without looking at each, press the CLR ALL soft
key.
9.30011
F1
F2
F3

Controlling System-Related Operations

~
V
F4
F5
Front-Panel Operation
caw062.eps
3
3-23