Retrieving The Factory Calibration - Hach POCKET COLORIMETER II Instruction Manual

Analysis systems
Hide thumbs Also See for POCKET COLORIMETER II:
Table of Contents

Advertisement

5. Press
READ/ENTER
Note: The minimum number of valid points is two. For example, if five points have been
entered, three can be deleted using this feature.
6. Press
to return to the measurement mode.
MENU

Retrieving the Factory Calibration

1. Select the range to restore factory default calibration. See
on page
2—4.
2. Press and hold the
3. Press the
ZERO/SCROLL
4. Press the
READ/ENTER
factory default calibration.
Note: For meters with factory-calibrated ranges or methods, Standard Calibration Adjust
(SCA) will be disabled when a user-entered method is programmed into the meter. To
turn SCA back on, restore the meter to factory default calibration.
User-Entered Calibration,
to delete. Repeat for all points to be deleted.
key until "USER", then "CAL" appears.
MENU
key to find dFL.
key to select dFL and restore the instrument to the
2—25