Phosphorus, Reactive,
For highly turbid or colored samples, add the contents of one Phosphate
Pretreatment Powder Pillow to 25 mL of sample. Mix well. Use this solution
(instead of the blank) to zero the instrument.
The following may interfere when present in concentrations exceeding those
:
listed below
Aluminum
Arsenate
Chromium
Copper
Hydrogen Sulfide
Iron
Nickel
pH, excess
buffering
continued
Table 3
Greater than 200 mg/L
Interferes at any level.
Greater than 100 mg/L
Greater than 10 mg/L
Interferes at any level
Greater than 100 mg/L
Greater than 300 mg/L
Highly buffered samples or extreme sample pH may exceed the
buffering capacity of the reagents and require sample
pretreatment. pH 2–10 is recommended.