Phosphonates

For water, wastewater, and seawater.
Persulfate UV Oxidation Method*
Hydrochloric Acid Solution (Cat. No. 884-49), followed by a distilled water
rinse. Do not clean glassware with commercial detergent.
Wear UV safety goggles while the UV lamp is on.
Do not handle the UV lamp surface. Fingerprints will etch the glass. Wipe the
lamp with a soft, clean tissue between samples.
The digestion in step 6 is normally completed in less than 10 minutes.
However, contaminated samples or a weak lamp can cause incomplete
phosphate conversion. Check conversion efficiency by running a longer
digestion and seeing if the readings increase.
Note:
The Pocket Colorimeter II is designed to measure solutions contained in sample cells.
dip the meter in the sample or pour the sample directly into the cell holder.
DO NOT
* Adapted from Blystone, P., Larson, P., A Rapid Method for Analysis of Phosphonate
Compounds, International Water Conference, Pittsburgh, PA, Oct. 26-29, 1981.
(0.1–2.5 to 1–125 mg/L)
1—39
Clean glassware with 1:1
Method 8007