Exiting The Calibration Routine; Deleting Calibration Points - Hach POCKET COLORIMETER II Instruction Manual

Analysis systems
Hide thumbs Also See for POCKET COLORIMETER II:
Table of Contents

Advertisement

User-Entered Calibration,
Note: When a factory calibration curve has been edited, the "calibration adjust" icon will
appear in the display.

Exiting the Calibration Routine

Exit the calibration routine by pressing the
mode. The instrument uses the last completed user-entered calibration or the
factory calibration if no user-entered calibration has been completed.

Deleting Calibration Points

1. Select the range containing user-entered calibration points. See
Ranges on page
2. Press and hold the
.
READ/ENTER
Note: Calibration points can also be deleted in Edit mode.
3. Press
ZERO/SCROLL
.
READ/ENTER
4. The left digit will flash. Press
appear after the numeral 9.)
continued
2—4.
key until "USER", then "CAL" appears. Press
MENU
to select the point to delete (e.g., S0 or S1 or S2). Press
ZERO/SCROLL
key to return to measurement
MENU
until "dEL" appears. ("dEL" will
2—24