Phosphonates,
6.
Insert the ultraviolet
lamp into the mixing bottle.
Turn the power supply on
and digest the sample for
10–15 minutes.
Phosphonates are
Note:
converted to orthophosphate
during digestion.
Caution: Wear UV safety
goggles while the lamp is on.
continued
7.
After digestion, turn the
power supply off and
remove the lamp.
Caution: Mixing bottle may
be hot if long digestion
periods are used.
1—42
8.
Fill a 10-mL sample cell
to the 10-mL mark with the
digested sample.
Cool the sample to
Note:
room temperature if it is hot
from digestion.